Contactez-nous

Chat en direct avec un représentant Tek. Service disponible de 9 h 00 à 17 h, CET.

Téléphone

Appelez-nous au 1-800-833-9200

Disponible de 9 h 00 à 17 h 00 CET jours ouvrables

Télécharger

Télécharger des manuels, des fiches techniques, des logiciels, etc. :

TYPE DE TÉLÉCHARGEMENT
MODÈLE OU MOT CLÉ

Feedback

Is the difference in the measured results between the TDSJIT3 software and using an infinite mode/Jitter mask acceptable?

Question :

Is the difference in the measured results between the TDSJIT3 software and using an infinite mode/Jitter mask acceptable?

Réponses :

Since longer acquisitions include drift and wander, they will normally be higher in jitter amplitude for a given population. So one must be careful and make sure the frequencies being measured are equivalent. So run TDSJIT3 in Free-Run mode for about 10 seconds to see what the longer term noise is - it should be similar to the WfmDB mask test.

Cette FAQ concerne :

Aucune série du produit

Produit :

Numéro de la FAQ 58496