Can the VM6000 software be put on a DPO70000 series oscilloscope? If not compatible+ what is the reason?

Question :

Can the VM6000 software be put on a DPO70000 series oscilloscope? If not compatible+ what is the reason?

Réponses :

No, the software cannot be used on the DPO70000 series. This series of oscilloscopes does not have the video trigger capability, or the ability to change between 50 Ohm input and 1 M Ohm input.  This function is essential to use of the VGA Measurement Interface Unit.